LONDON (TIP): Researchers have developed a new batteryless cardiac pacemaker which is based on an automatic wristwatch and is powered by heart motion. “Batteries are a limiting factor in today’s medical implants. Once they reach a critically low energy level, physicians see themselves forced to replace a correctly functioning medical device in a surgical intervention,” said Adrian Zurbuchen from the University of Bern, Switzerland. “This is an unpleasant scenario which increases costs and the risk of complications for patients,” Zurbuchen said. He has now come up with a way to power a cardiac pacemaker with an alternative energy source – the heart motion.
Four years ago Professor Rolf Vogel, a cardiologist and engineer at the University of Bern, had the idea of using an automatic wristwatch mechanism to harvest the energy of heart motion. “The heart seems to be a very promising energy source because its contractions are repetitive and present for 24 hours a day, 7 days a week,” Zurbuchen said.
“Furthermore, the automatic clockwork, invented in 1777, has a good reputation as a reliable technology to scavenge energy from motion.” The researchers’ first prototype is based on a commercially available automatic wristwatch. All unnecessary pa8rts were removed to reduce weight and size. They developed a custommade ho8using with eyelets that allows suturing the device directly onto the myocardium. The prototype works the same way it would on a person’s wrist.
When it is exposed to an external acceleration, the eccentric mass of the clockwork starts rotating. This progressively winds a mechanical spring. After the spring is fully charged it unwinds and thereby spins an electrical micro-generator. To test the prototype, the researchers developed an electronic circuit to transform and store the signal into a small buffer capacity. They then connected the system to a custom-made cardiac pacemaker.
The system worked in three steps. First, the harvesting prototype acquired energy from the heart. Second, the energy was temporarily stored in the buffer capacity. And finally, the buffered energy was used by the pacemaker to apply minute st8imuli to the heart. The researchers successfully tested the system in in vivo experiments with domestic pigs. The mechanism allowed them for the first time to perform batteryless overdrive-pacing at 130 beats per minute.
“We have shown that it is possible to pace the heart using the power of its own motionThe next step in our prototype is to integrate both the electronic circuit for energy storage and the custom-made pacemaker directly into the harvesting device. This will eliminate the need for leads,” Zurbuchen said.